Displaying Data
6
6000 Series Oscilloscope User’s Guide
289
When the
FlexRay
mode is selected, you can use the
Settings
softkey to access the BUSDOCTOR Menu for changing the
BUSDOCTOR 2 control/operating modes.
4
If the decode line does not appear on the display, press the
Decode
softkey to turn it on. BUSDOCTOR must be connected
and running in asynchronous or synchronous mode for a
meaningful display.
5
If the oscilloscope is stopped, press the
Run/Stop
key to
acquire and decode data.
You can use the
Delayed
horizontal sweep mode for easier
navigation of the decoded data.
N O T E
Changing the BUSDOCTOR settings in the Serial Decode Menu also
changes them in the FlexRay Trigger Menu.