6000 Series Oscilloscope User’s Guide
4
Triggering the Oscilloscope
4
Press the
BUSDOCTOR 2
softkey to access the BUSDOCTOR 2
Menu.
Specifying the BUSDOCTOR LAN Address
This is necessary when the oscilloscope controls BUSDOCTOR
(but not when a PC with VISION software controls
BUSDOCTOR).
1
From the BUSDOCTOR 2 Menu, press the
LAN Address
softkey
to access the BUSDOCTOR LAN Addresses Menu.
2
Enter the BUSDOCTOR’s 4-byte IP address:
a
Press the
Value
softkey and use the Entry knob to select a
byte value.
b
Press the
Next
softkey to access the next byte.
3
When you have finished entering the IP address, press the
Apply
softkey.
This is sufficient for connection via a hub or switch.
If the oscilloscope is connected directly to the BUSDOCTOR (in
a point-to-point configuration), the following additional steps
must be taken.
Baud Rate
Mode
Channel
FlexRay
trigger
Return to
previous menu
TDMA Struct
(Sync Mode)
BUSDOCTOR
LAN Address
Summary of Contents for InfiniiVision DSO6014L
Page 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Page 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Page 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Page 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Page 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Page 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Page 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Page 370: ......