6000 Series Oscilloscope User’s Guide
4
Triggering the Oscilloscope
•
HF
Reject
adds a 50 kHz low-pass filter in the trigger path
to remove high frequency components from the trigger
waveform. You can use HF Reject to remove
high-frequency noise, such as AM or FM broadcast stations
or noise from fast system clocks, from the trigger path.
To set Holdoff
1
Press the
Mode/Coupling
key.
2
Turn the Entry knob
to increase or decrease the trigger
holdoff time shown in the
Holdoff
softkey.
Holdoff sets the amount of time that the oscilloscope waits
before re-arming the trigger circuitry. Use Holdoff to stabilize
the display of complex waveforms.
To get a stable trigger on the pulse burst shown below, set the
holdoff time to be >200 ns but <600 ns.
By setting the Holdoff, you can synchronize triggers. The
oscilloscope will trigger on one edge of the waveform, and
ignore further edges until the holdoff time expires. The
oscilloscope will then re-arm the trigger circuit to search for the
next edge trigger. This allows the oscilloscope to trigger on a
repeating pattern in a waveform.
Scope triggers here
Holdoff
200 ns
600 ns
Summary of Contents for InfiniiVision DSO6014L
Page 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Page 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Page 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Page 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Page 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Page 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Page 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Page 370: ......