Displaying Data
6
6000 Series Oscilloscope User’s Guide
273
You can use the
Delayed
horizontal sweep mode for easier
navigation of the acquired data.
Interpreting Decoded SPI Data
•
White lines show an idle bus.
•
Blue rectangles show an active bus (inside a packet/frame).
•
The number of clocks in a frame appears in light-blue above
the frame, to the right.
•
Decoded hexadecimal data values appear in white and may
extend beyond the frame if an idle period follows.
•
Red dots in the decode line indicate that there is data that is
not being displayed. Scroll or expand the horizontal scale to
view the information.
Summary of Contents for InfiniiVision DSO6014L