Agilent E5250A User’s Guide, Edition 11
7-13
Command Reference
*STB?
*STB?
This query command reads the Status Byte Register (reads Master Summary Status
bit, not Request for Service Message).
For bit6, this command reads MSS, not Request for Service (RQS). Refer to “Status
Reporting Structure” on page 7-48.
Syntax
*STB?
Query response
register
<newline><^END>
Semantics
The following table shows the bits of the Status Byte Register, and the
binary-weighted
decimal
value of each bit.
Example
OUTPUT @Hp5250;"*STB?"
ENTER @Hp5250;A
Parameter Explanation
register
decimal integer value that is the sum of the binary-weighted values
for the set bits
bit
binary-
weight
description
0
1
not used. always 0.
1
2
not used. always 0.
2
4
not used. always 0.
3
8
not used. always 0.
4
16
MAV (Message Available summary-message)
5
32
ESB (Event Status Bit)
6
64
MSS (Master Summary Status)
7
128
not used. always 0.
Summary of Contents for Agilent E5250A
Page 1: ...Agilent Technologies Agilent E5250A Low Leakage Switch Mainframe User s Guide ...
Page 10: ......
Page 16: ...Contents 6 Agilent E5250A User s Guide Edition 11 Contents 12 SCPI Command Summary ...
Page 17: ...1 Introduction ...
Page 32: ...1 16 Agilent E5250A User s Guide Edition 11 Introduction Options and Accessories ...
Page 33: ...2 Installation ...
Page 51: ...3 Executing Self Test and Leak Test ...
Page 71: ...4 Setting up Measurement Environment ...
Page 95: ...5 Controlling the E5250A ...
Page 137: ...6 Programming the E5250A ...
Page 169: ...7 Command Reference ...
Page 224: ...7 56 Agilent E5250A User s Guide Edition 11 Command Reference Output Queue ...
Page 225: ...8 VXIplug play Driver ...
Page 256: ...8 32 Agilent E5250A User s Guide Edition 11 VXIplug play Driver Driver Functions ...
Page 257: ...9 Executing Sample Programs ...
Page 295: ...10 Specifications ...
Page 305: ...11 Error Messages ...
Page 318: ...11 14 Agilent E5250A User s Guide Edition 11 Error Messages E5250A Specific Error Messages ...
Page 319: ...12 SCPI Command Summary ...