Agilent E5250A User’s Guide, Edition 11
6-29
Programming the E5250A
Using the Capacitance Compensation Routine
Capacitance Compensation Routine
This routine is an subprogram used to compensate for capacitance measurement
error caused by the E5250A with the E5252A matrix card installed. The subprogram
is used with Agilent 4284A Precision LCR Meter. You enter capacitance and
conductance data measured by the 4284A as a parameter, and this subprogram
returns the compensated value. This routine runs on HP BASIC.
Syntax:
Ccompen_5250(Freq,L1,L2_h,L2_l,L3_h,L3_l,C_m,G_m,C_c,G_c)
Parameter:
Freq
Measurement Frequency (Hz): 1000 Hz to 1000000 Hz (1 kHz to
1 MHz).
L1
Length (in meters) of 16494A triaxial cable connected from E5252A
output to connector plate.
L2_h
Length (in meters) of triaxial cable connected from connector plate to
DUT (
High
terminal of capacitor).
If you do not use triaxial cable, enter
0
for this parameter.
L2_l
Length (in meters) of triaxial cable connected from connector plate to
DUT (
Low
terminal of capacitor).
If you do not use triaxial cable, enter
0
for this parameter.
L3_h
Length (in meters) of coaxial cable connected from connector plate to
DUT (
High
terminal of capacitor).
If you do not use coaxial cable, enter
0
for this parameter.
L3_l
Length (in meters) of coaxial cable connected from connector plate to
DUT (
Low
terminal of capacitor).
If you do not use coaxial cable, enter
0
for this parameter.
C_m
Capacitance data (in F) measured by 4284A.
G_m
Conductance data (in S) measured by 4284A.
C_c
Compensated capacitance value (in F) returned by Capacitance
Compensation Routine.
G_c
Compensated conductance value (in S) returned by Capacitance
Compensation Routine.
Summary of Contents for Agilent E5250A
Page 1: ...Agilent Technologies Agilent E5250A Low Leakage Switch Mainframe User s Guide ...
Page 10: ......
Page 16: ...Contents 6 Agilent E5250A User s Guide Edition 11 Contents 12 SCPI Command Summary ...
Page 17: ...1 Introduction ...
Page 32: ...1 16 Agilent E5250A User s Guide Edition 11 Introduction Options and Accessories ...
Page 33: ...2 Installation ...
Page 51: ...3 Executing Self Test and Leak Test ...
Page 71: ...4 Setting up Measurement Environment ...
Page 95: ...5 Controlling the E5250A ...
Page 137: ...6 Programming the E5250A ...
Page 169: ...7 Command Reference ...
Page 224: ...7 56 Agilent E5250A User s Guide Edition 11 Command Reference Output Queue ...
Page 225: ...8 VXIplug play Driver ...
Page 256: ...8 32 Agilent E5250A User s Guide Edition 11 VXIplug play Driver Driver Functions ...
Page 257: ...9 Executing Sample Programs ...
Page 295: ...10 Specifications ...
Page 305: ...11 Error Messages ...
Page 318: ...11 14 Agilent E5250A User s Guide Edition 11 Error Messages E5250A Specific Error Messages ...
Page 319: ...12 SCPI Command Summary ...