10-6
Agilent E5250A User’s Guide, Edition 11
Specifications
Option 501 Multiplexer
Option 501 Multiplexer
Switch Specifications
This specification is defined with the module plugged into the mainframe. The value
depends on the following temperature and humidity conditions.
•
A
: 23
°
C ± 5
°
C, 5 % to 60 % RH
•
B
: 5
°
C to 18
°
C, 28
°
C to 40
°
C, 5 % to 60 % RH
•
C
: 28
°
C to 40
°
C, 60 % to 80 % RH
A B C
Max. Current Rating (A)
1.0
1.0
1.0
Max. Voltage Rating (V)
Channel to Guard
200
200
200
Channel to Channel
300
300
300
Closed Channel Residual Resistance (
Ω
)
Low Leakage I-V Port
0.6
0.6
0.6
BIAS Port
a
1.0
1.0
1.0
Channel Isolation Resistance (
Ω
)
Low Leakage I-V Port
10
13
5
×
10
12
2
×
10
12
BIAS Port
10
9
5
×
10
8
2
×
10
8
a. When 0
Ω
resistors are mounted on the card as the protective resis-
tor.
Summary of Contents for Agilent E5250A
Page 1: ...Agilent Technologies Agilent E5250A Low Leakage Switch Mainframe User s Guide ...
Page 10: ......
Page 16: ...Contents 6 Agilent E5250A User s Guide Edition 11 Contents 12 SCPI Command Summary ...
Page 17: ...1 Introduction ...
Page 32: ...1 16 Agilent E5250A User s Guide Edition 11 Introduction Options and Accessories ...
Page 33: ...2 Installation ...
Page 51: ...3 Executing Self Test and Leak Test ...
Page 71: ...4 Setting up Measurement Environment ...
Page 95: ...5 Controlling the E5250A ...
Page 137: ...6 Programming the E5250A ...
Page 169: ...7 Command Reference ...
Page 224: ...7 56 Agilent E5250A User s Guide Edition 11 Command Reference Output Queue ...
Page 225: ...8 VXIplug play Driver ...
Page 256: ...8 32 Agilent E5250A User s Guide Edition 11 VXIplug play Driver Driver Functions ...
Page 257: ...9 Executing Sample Programs ...
Page 295: ...10 Specifications ...
Page 305: ...11 Error Messages ...
Page 318: ...11 14 Agilent E5250A User s Guide Edition 11 Error Messages E5250A Specific Error Messages ...
Page 319: ...12 SCPI Command Summary ...