Agilent E5250A User’s Guide, Edition 11
3-15
Executing Self-Test and Leak Test
Using the Self-Test Utility
To Execute Leak Test
NOTE
Leak Test is for checking leakage current of the E5250A with plug-in cards
installed. If leak test result is too large for your application, contact nearest Agilent
Technologies Service Center. Leak test does not verify specifications of the E5250A
and installed cards.
The Leak Test applies 10 Vdc to the specified input port of the E5250A, and
measures the leakage current. The test limits are shown in Table 3-4.
You specify an input port of the E5250A, and specify the 4155/4156 SMU to use.
The Leak Test automatically connects the specified input port to each output port
consecutively, and uses the 4155/4156 SMU to measure the leakage current for each
channel.
You need to execute this test for each input port.
Table 3-4
Test Limits of Leak Test
Card Instrument Connection
Input
Port Test
Limit
E5252A
4155
non-Kelvin
IV1, IV2
±4.12 pA
IV3, IV4, IV5, IV6
±1.01 nA
4156
non-Kelvin
IV1, IV2
±1.14 pA
IV3, IV4, IV5, IV6
±1.01 nA
Kelvin
IV1, IV2
±2.24 pA
IV3, IV4, IV5, IV6
±2.02 nA
E5255A
4155
non-Kelvin
IV1 to IV6
±4.12 pA
4156
non-Kelvin
IV1 to IV6
±1.14 pA
Summary of Contents for Agilent E5250A
Page 1: ...Agilent Technologies Agilent E5250A Low Leakage Switch Mainframe User s Guide ...
Page 10: ......
Page 16: ...Contents 6 Agilent E5250A User s Guide Edition 11 Contents 12 SCPI Command Summary ...
Page 17: ...1 Introduction ...
Page 32: ...1 16 Agilent E5250A User s Guide Edition 11 Introduction Options and Accessories ...
Page 33: ...2 Installation ...
Page 51: ...3 Executing Self Test and Leak Test ...
Page 71: ...4 Setting up Measurement Environment ...
Page 95: ...5 Controlling the E5250A ...
Page 137: ...6 Programming the E5250A ...
Page 169: ...7 Command Reference ...
Page 224: ...7 56 Agilent E5250A User s Guide Edition 11 Command Reference Output Queue ...
Page 225: ...8 VXIplug play Driver ...
Page 256: ...8 32 Agilent E5250A User s Guide Edition 11 VXIplug play Driver Driver Functions ...
Page 257: ...9 Executing Sample Programs ...
Page 295: ...10 Specifications ...
Page 305: ...11 Error Messages ...
Page 318: ...11 14 Agilent E5250A User s Guide Edition 11 Error Messages E5250A Specific Error Messages ...
Page 319: ...12 SCPI Command Summary ...