Agilent E5250A User’s Guide, Edition 11
7-27
Command Reference
[:ROUTe]:BIAS:PORT
[:ROUTe]:BIAS:PORT
NOTE
This command applies only to the E5252A card.
For the E5255A card, the input Bias Port cannot be selected, and is always the
built-in Bias Port.
The Bias Port and Couple Port cannot be set to same port number.
This command selects which input port is the Bias Port on the specified card. For
each card, you can specify the same or different Bias Port. For details, refer to "Bias
Mode" in Chapter 5.
At *RST, Bias Port is set to 10 for the E5252A.
Syntax
[:ROUTe]:BIAS:PORT
card_number
,
bias_port
Example
OUTPUT @Hp5250;":ROUT:BIAS:PORT ALL,6"
Parameter Explanation
card_number
card number:
For Auto Config mode:
0
or
ALL
For Normal Config mode:
1
,
2
,
3
,
4
, or
ALL
bias_port
Input port number:
1
to
10
.
Summary of Contents for Agilent E5250A
Page 1: ...Agilent Technologies Agilent E5250A Low Leakage Switch Mainframe User s Guide ...
Page 10: ......
Page 16: ...Contents 6 Agilent E5250A User s Guide Edition 11 Contents 12 SCPI Command Summary ...
Page 17: ...1 Introduction ...
Page 32: ...1 16 Agilent E5250A User s Guide Edition 11 Introduction Options and Accessories ...
Page 33: ...2 Installation ...
Page 51: ...3 Executing Self Test and Leak Test ...
Page 71: ...4 Setting up Measurement Environment ...
Page 95: ...5 Controlling the E5250A ...
Page 137: ...6 Programming the E5250A ...
Page 169: ...7 Command Reference ...
Page 224: ...7 56 Agilent E5250A User s Guide Edition 11 Command Reference Output Queue ...
Page 225: ...8 VXIplug play Driver ...
Page 256: ...8 32 Agilent E5250A User s Guide Edition 11 VXIplug play Driver Driver Functions ...
Page 257: ...9 Executing Sample Programs ...
Page 295: ...10 Specifications ...
Page 305: ...11 Error Messages ...
Page 318: ...11 14 Agilent E5250A User s Guide Edition 11 Error Messages E5250A Specific Error Messages ...
Page 319: ...12 SCPI Command Summary ...