9-6
Agilent E5250A User’s Guide, Edition 11
Executing Sample Programs
Vth and Capacitance Measurement Program
Display Setup:
Also the DATA VARIAVLES fields set VTH and BETA.
Auto Analysis Setup:
*LINE1: [ TANGENT
] line on [ Y1 ]
at a point where
[ GRAD
] = [ MAX(GRAD)
]
-----------------------------------------------------------
*LINE2: [
]
-----------------------------------------------------------
*MARKER:
At a point where
[ GRAD
] = [ MAX(GRAD)
]
[
]
---------------------
*Interpolate: [OFF]
Capacitance Measurement Setup
Measurement setup of the E5250A and 4284A for this program is as follows:
•
E5250A/E5252A Setup:
•
4284A Setup:
Xaxis Y1axis Y2axis
NAME
Vg
SQRTId
GRAD
SCALE
LINEAR
LINEAR
LINEAR
MIN
0 V
0
0
MAX
5 V
200 m
80 m
Slot No
Channels
1
10905 and 11006
2
not used.
3
not used.
4
not used.
Measurement Frequency
100 kHz
Signal Level
0.03 V
Measurement Function
Cp-G
Summary of Contents for Agilent E5250A
Page 1: ...Agilent Technologies Agilent E5250A Low Leakage Switch Mainframe User s Guide ...
Page 10: ......
Page 16: ...Contents 6 Agilent E5250A User s Guide Edition 11 Contents 12 SCPI Command Summary ...
Page 17: ...1 Introduction ...
Page 32: ...1 16 Agilent E5250A User s Guide Edition 11 Introduction Options and Accessories ...
Page 33: ...2 Installation ...
Page 51: ...3 Executing Self Test and Leak Test ...
Page 71: ...4 Setting up Measurement Environment ...
Page 95: ...5 Controlling the E5250A ...
Page 137: ...6 Programming the E5250A ...
Page 169: ...7 Command Reference ...
Page 224: ...7 56 Agilent E5250A User s Guide Edition 11 Command Reference Output Queue ...
Page 225: ...8 VXIplug play Driver ...
Page 256: ...8 32 Agilent E5250A User s Guide Edition 11 VXIplug play Driver Driver Functions ...
Page 257: ...9 Executing Sample Programs ...
Page 295: ...10 Specifications ...
Page 305: ...11 Error Messages ...
Page 318: ...11 14 Agilent E5250A User s Guide Edition 11 Error Messages E5250A Specific Error Messages ...
Page 319: ...12 SCPI Command Summary ...