Agilent E5250A User’s Guide, Edition 11
8-29
VXIplug&play Driver
Driver Functions
hpe5250a_selectCompenFile
This function specifies capacitance/conductance compensation data file used to
compensate C/G by using hpe5250a_compenC.
Syntax
ViStatus _VI_FUNC hpe5250a_selectCompenFile(ViSession vi,
ViString file_name);
Parameters
vi
Instrument handle returned from hpe5250a_init( ).
file_name
Compensation data file name. Use absolute path. If the value is
NULL string, the default data is used.
Remarks
If you change the compensation data, copy the default data shown below, and
modify the data for your measurement cable. You will need to change the data for
DATA05 and 06, and/or DATA07 and 08 corresponding to your cables. To measure
and change the compensation data, refer to Chapter 6. To get the R, L, and C value,
measure R, L, and C of the cable using the 4284A, and divide them by cable length
(in m). Compensation data must be the value for 1 m length. Do not change the data
format in the file.
# E5250A C Compensation coefficient data table
#
# CAUTION : Do not add or delete "REVISION" line and "DATAxx" line.
# Change the value for R,L,C of DATA05,06,07 or 08.
#
REVISION
A.03.00
#
R [ohm]
L [H]
C [F]
DATA00
74.65E-3
140.00E-9
58.44E-12
# Frame Path 1
DATA01
75.41E-3
90.00E-9
67.13E-12
# Frame Path 2
DATA02
231.41E-3
450.00E-9
178.85E-12
# Card Path High
DATA03
177.56E-3
390.00E-9
135.45E-12
# Card Path Low
DATA04
100.70E-3
400.00E-9
80.00E-12
# Triax Cable [/m]
DATA05
100.70E-3
400.00E-9
80.00E-12
# User Triax Cbl H [/m]
DATA06
100.70E-3
400.00E-9
80.00E-12
# User Triax Cbl L [/m]
DATA07
114.00E-3
544.00E-9
130.00E-12
# User Coax Cbl H [/m]
DATA08
114.00E-3
544.00E-9
130.00E-12
# User Coax Cbl L [/m]
DATA09
0.00E-3
0.00E-9
1.20E-12
# Stray Capacitance
# END of Data
Summary of Contents for Agilent E5250A
Page 1: ...Agilent Technologies Agilent E5250A Low Leakage Switch Mainframe User s Guide ...
Page 10: ......
Page 16: ...Contents 6 Agilent E5250A User s Guide Edition 11 Contents 12 SCPI Command Summary ...
Page 17: ...1 Introduction ...
Page 32: ...1 16 Agilent E5250A User s Guide Edition 11 Introduction Options and Accessories ...
Page 33: ...2 Installation ...
Page 51: ...3 Executing Self Test and Leak Test ...
Page 71: ...4 Setting up Measurement Environment ...
Page 95: ...5 Controlling the E5250A ...
Page 137: ...6 Programming the E5250A ...
Page 169: ...7 Command Reference ...
Page 224: ...7 56 Agilent E5250A User s Guide Edition 11 Command Reference Output Queue ...
Page 225: ...8 VXIplug play Driver ...
Page 256: ...8 32 Agilent E5250A User s Guide Edition 11 VXIplug play Driver Driver Functions ...
Page 257: ...9 Executing Sample Programs ...
Page 295: ...10 Specifications ...
Page 305: ...11 Error Messages ...
Page 318: ...11 14 Agilent E5250A User s Guide Edition 11 Error Messages E5250A Specific Error Messages ...
Page 319: ...12 SCPI Command Summary ...