Agilent E5250A User’s Guide, Edition 11
8-11
VXIplug&play Driver
Driver Functions
hpe5250a_closeList
This function will connect the input ports to the output ports specified by the
channel list.
The parameter ’closechan_list’ is an array of integers with each integer representing
one channel. The last number of the ’closechan_list’ should be "0" (numeric zero)
to identify the end of the list. The maximum number of channels that can be
specified by the list is 100.
Syntax
ViStatus _VI_FUNC hpe5250a_closeList(ViSession vi,
ViInt32_VI_FAR closechan_list[ ] );
Parameters
vi
Instrument handle returned from hpe5250a_init( ).
closechan_list[ ]
Channel numbers to connect. 5 digits integer. ABCDE. where
A: card number, BC: input port number, DE: output port
number. Top zero(s) can be ignored. For example, if A=0,
BC=01, and DE=01, channel number should be 101 instead of
00101.
Summary of Contents for Agilent E5250A
Page 1: ...Agilent Technologies Agilent E5250A Low Leakage Switch Mainframe User s Guide ...
Page 10: ......
Page 16: ...Contents 6 Agilent E5250A User s Guide Edition 11 Contents 12 SCPI Command Summary ...
Page 17: ...1 Introduction ...
Page 32: ...1 16 Agilent E5250A User s Guide Edition 11 Introduction Options and Accessories ...
Page 33: ...2 Installation ...
Page 51: ...3 Executing Self Test and Leak Test ...
Page 71: ...4 Setting up Measurement Environment ...
Page 95: ...5 Controlling the E5250A ...
Page 137: ...6 Programming the E5250A ...
Page 169: ...7 Command Reference ...
Page 224: ...7 56 Agilent E5250A User s Guide Edition 11 Command Reference Output Queue ...
Page 225: ...8 VXIplug play Driver ...
Page 256: ...8 32 Agilent E5250A User s Guide Edition 11 VXIplug play Driver Driver Functions ...
Page 257: ...9 Executing Sample Programs ...
Page 295: ...10 Specifications ...
Page 305: ...11 Error Messages ...
Page 318: ...11 14 Agilent E5250A User s Guide Edition 11 Error Messages E5250A Specific Error Messages ...
Page 319: ...12 SCPI Command Summary ...