Agilent E5250A User’s Guide, Edition 11
9-21
Executing Sample Programs
HCI Measurement Program
Step 5. Applies stress and characterizes the parameters
This step performs the following:
1. Applies the stress,
Vdstr
and
Vgstr
, to the valid devices.
Before running the
SP_MUX_M
program, you must determine the
Vdstr
(drain
stress voltage) by Id-Vds measurement (using setup file
IDVD.MES
), then edit
the
SP_MUX_M
program to input value of
Vdstr
. Refer to “Executing the
Program” on page 9-27.
2. Performs the interim characterization. This sets up the 4155/4156 by using the
PARAM.MES
setup file, then measures the parameters (
Idlin
,
Gmmax
,
Vtext
and
Vtci
) for each device.
3. Judges whether to terminate the stress/interim characterization cycle. This cycle
is terminated when total stress time reaches 100,000 sec or when there is no
device satisfies all of the following conditions:
•
Idlin
value is within ±10 % of initial value.
•
Gmmax
value is within ±10 % of initial value.
•
Vtext
value is within ±20 mV of initial value.
•
Vtci
value is within ±20 mV of initial value.
If none of these termination conditions has occurred, stress time (duration) is
increased, and these 3 steps are repeated. The default
cumulative
stress times are
defined in the program as follows. To change these
cumulative
stress times, refer
to “Modifying the Program” on page 9-35.
Stress
cycle
Stress time
Cumulative
stress time
Stress
cycle
Stress time
Cumulative
stress time
1
10 sec
10 sec
8
1000 sec
2000 sec
2
10 sec
20 sec
9
3000 sec
5000 sec
3
30 sec
50 sec
10
5000 sec
10000 sec
4
50 sec
100 sec
11
10000 sec
20000 sec
5
100 sec
200 sec
12
30000 sec
50000 sec
6
300 sec
500 sec
13
50000 sec
100000 sec
7
500 sec
1000 sec
Summary of Contents for Agilent E5250A
Page 1: ...Agilent Technologies Agilent E5250A Low Leakage Switch Mainframe User s Guide ...
Page 10: ......
Page 16: ...Contents 6 Agilent E5250A User s Guide Edition 11 Contents 12 SCPI Command Summary ...
Page 17: ...1 Introduction ...
Page 32: ...1 16 Agilent E5250A User s Guide Edition 11 Introduction Options and Accessories ...
Page 33: ...2 Installation ...
Page 51: ...3 Executing Self Test and Leak Test ...
Page 71: ...4 Setting up Measurement Environment ...
Page 95: ...5 Controlling the E5250A ...
Page 137: ...6 Programming the E5250A ...
Page 169: ...7 Command Reference ...
Page 224: ...7 56 Agilent E5250A User s Guide Edition 11 Command Reference Output Queue ...
Page 225: ...8 VXIplug play Driver ...
Page 256: ...8 32 Agilent E5250A User s Guide Edition 11 VXIplug play Driver Driver Functions ...
Page 257: ...9 Executing Sample Programs ...
Page 295: ...10 Specifications ...
Page 305: ...11 Error Messages ...
Page 318: ...11 14 Agilent E5250A User s Guide Edition 11 Error Messages E5250A Specific Error Messages ...
Page 319: ...12 SCPI Command Summary ...