Agilent E5250A User’s Guide, Edition 11
4-19
Setting up Measurement Environment
Connecting the E5250A Input
Figure 4-7
Connection of Triax-Dual Triax Adapter
By this connection, input signal for SMU INPUT connectors is as follows:
If you make non-Kelvin connection for the E5250A outputs, use the odd input port
only. Because non-Kelvin connection uses only Force signal for measurement.
SMU INPUT
AUX
To
SMU
1
2
3
4
5
6
Agilent E5250A
From
GND
TRIAX-DUAL TRIAX
Adapter
Agilent 16493H Cable
To
SMU INPUT
To
GNDU
SMU INPUT Port Number
Input Signal
1, 3, or 5
Force
2, 4, or 6
Sense
Summary of Contents for Agilent E5250A
Page 1: ...Agilent Technologies Agilent E5250A Low Leakage Switch Mainframe User s Guide ...
Page 10: ......
Page 16: ...Contents 6 Agilent E5250A User s Guide Edition 11 Contents 12 SCPI Command Summary ...
Page 17: ...1 Introduction ...
Page 32: ...1 16 Agilent E5250A User s Guide Edition 11 Introduction Options and Accessories ...
Page 33: ...2 Installation ...
Page 51: ...3 Executing Self Test and Leak Test ...
Page 71: ...4 Setting up Measurement Environment ...
Page 95: ...5 Controlling the E5250A ...
Page 137: ...6 Programming the E5250A ...
Page 169: ...7 Command Reference ...
Page 224: ...7 56 Agilent E5250A User s Guide Edition 11 Command Reference Output Queue ...
Page 225: ...8 VXIplug play Driver ...
Page 256: ...8 32 Agilent E5250A User s Guide Edition 11 VXIplug play Driver Driver Functions ...
Page 257: ...9 Executing Sample Programs ...
Page 295: ...10 Specifications ...
Page 305: ...11 Error Messages ...
Page 318: ...11 14 Agilent E5250A User s Guide Edition 11 Error Messages E5250A Specific Error Messages ...
Page 319: ...12 SCPI Command Summary ...