Agilent E5250A User’s Guide, Edition 11
3-19
Executing Self-Test and Leak Test
Using the Self-Test Utility
8. Connect the cable from the 4155/4156 SMU port specified by the SMUm
softkey (
m
=1 to 6) in step 5 to the E5250A input port specified by the IVn
softkey (
n
=1 to 6) in step 4.
•
Use Kelvin triaxial cable for Kelvin connection
•
Use triaxial cable for non-Kelvin connection
9. Press the CONTINUE softkey to execute the Leak Test.
After the Leak Test is completed for the input port specified by the IVn softkey,
the test result (PASS or FAIL) is displayed on the 4155/4156 screen. And the
following softkeys are displayed:
To display the leak current measurement data on the 4155/4156 screen, press the
REPORT CRT softkey.
10. To return to the LEAK TEST MENU, select the LEAK TEST softkey. And to
perform the Leak Test for another input port, repeat steps 3 to 9.
Softkey Description
REPORT CRT
Displays the Leak Test results on the 4155/4156 screen.
LEAK TEST
Returns to the LEAK TEST MENU.
Summary of Contents for Agilent E5250A
Page 1: ...Agilent Technologies Agilent E5250A Low Leakage Switch Mainframe User s Guide ...
Page 10: ......
Page 16: ...Contents 6 Agilent E5250A User s Guide Edition 11 Contents 12 SCPI Command Summary ...
Page 17: ...1 Introduction ...
Page 32: ...1 16 Agilent E5250A User s Guide Edition 11 Introduction Options and Accessories ...
Page 33: ...2 Installation ...
Page 51: ...3 Executing Self Test and Leak Test ...
Page 71: ...4 Setting up Measurement Environment ...
Page 95: ...5 Controlling the E5250A ...
Page 137: ...6 Programming the E5250A ...
Page 169: ...7 Command Reference ...
Page 224: ...7 56 Agilent E5250A User s Guide Edition 11 Command Reference Output Queue ...
Page 225: ...8 VXIplug play Driver ...
Page 256: ...8 32 Agilent E5250A User s Guide Edition 11 VXIplug play Driver Driver Functions ...
Page 257: ...9 Executing Sample Programs ...
Page 295: ...10 Specifications ...
Page 305: ...11 Error Messages ...
Page 318: ...11 14 Agilent E5250A User s Guide Edition 11 Error Messages E5250A Specific Error Messages ...
Page 319: ...12 SCPI Command Summary ...