Agilent E5250A User’s Guide, Edition 11
9-27
Executing Sample Programs
HCI Measurement Program
Executing the Program
You should have already added the bias source control routine and set up the
measurement environment as described previously.
Before running the program, do following:
1. Set the GPIB address of instruments as follows:
•
4155/4156: 717
•
E5250A: 722
•
Bias Source: Set the GPIB addresses to match your control routine
(Address 701 to 703 are defaults used in
SP_MUX_M
program)
2. Turn on all instruments and the computer.
3. Boot up HP BASIC operating system.
4. Insert a diskette into the computer's disk drive. The diskette must contain the
HCI test program (SP_MUX in this example) you modified in “Adding the Bias
Source Control Routine” on page 9-23.
5. Specify the computer's mass storage device (disk drive) the diskette is inserted
by using MSI statement. For example, if the device is ":CS80,702,1", enter the
following command:
MSI ":CS80,702,1"
6. Load your HCI test program. The following example loads and displays
SP_MUX program, in this example.
LOAD "SP_MUX"
EDIT
7. Insert the E5250A Program Disk into the 4155/4156's disk drive, then get the
setup file
IDVD.MES
to set up the 4155/4156.
8. Enter the following command from the command line of the computer:
OUTPUT 722;":ROUT:CLOS (@10101,20101,30101)"
9. Press Single key on the 4155/4156's front panel. The Id-Vds measurement is
performed and the Id-Vds measurement result is displayed as shown in Figure
9-5.
10. Determine the drain stress voltage (
Vdstr
) from the above curve.
Recommended maximum value is about 0.5 V below actual breakdown. Then,
edit the HCI test program (SP_MUX in this example) , and define the
Vdstr
value. The value should be defined in the following program line.
Summary of Contents for Agilent E5250A
Page 1: ...Agilent Technologies Agilent E5250A Low Leakage Switch Mainframe User s Guide ...
Page 10: ......
Page 16: ...Contents 6 Agilent E5250A User s Guide Edition 11 Contents 12 SCPI Command Summary ...
Page 17: ...1 Introduction ...
Page 32: ...1 16 Agilent E5250A User s Guide Edition 11 Introduction Options and Accessories ...
Page 33: ...2 Installation ...
Page 51: ...3 Executing Self Test and Leak Test ...
Page 71: ...4 Setting up Measurement Environment ...
Page 95: ...5 Controlling the E5250A ...
Page 137: ...6 Programming the E5250A ...
Page 169: ...7 Command Reference ...
Page 224: ...7 56 Agilent E5250A User s Guide Edition 11 Command Reference Output Queue ...
Page 225: ...8 VXIplug play Driver ...
Page 256: ...8 32 Agilent E5250A User s Guide Edition 11 VXIplug play Driver Driver Functions ...
Page 257: ...9 Executing Sample Programs ...
Page 295: ...10 Specifications ...
Page 305: ...11 Error Messages ...
Page 318: ...11 14 Agilent E5250A User s Guide Edition 11 Error Messages E5250A Specific Error Messages ...
Page 319: ...12 SCPI Command Summary ...