Continued from previous page...
1
Reverse the cable connections to the calibrator to create a -10 Vdc
voltage standard. You
must
physically reverse the cables.
DO NOT
switch the output polarity of the calibrator.
2
Perform a –10V DC gain calibration. Press
to enter the
calibration menu. Press
again to begin the adjustment
procedure. Be sure to allow time for thermal offsets to stabilize
(usually about 1 minute).
3
Use the knob
and
to set the number in the display to the
difference
of the calibrator output and the measured offset (from step 3)
and press
. Using the previous example values, enter 10
µ
V
minus 10.001 volts or –10.000990 volts.
4
When the adjustment finishes, verify that new readings fall
within
±
30
µ
V of the calibrator output minus the offset.
Chapter 4 Calibration Procedures
–10 Vdc Adjustment Procedure (Optional)
86
Summary of Contents for 34970A
Page 1: ...Agilent 34970A Data Acquisition Switch Unit Service Guide ...
Page 34: ...32 ...
Page 35: ...2 Quick Start 2 ...
Page 49: ...3 Front Panel Overview 3 ...
Page 62: ...60 ...
Page 63: ...4 Calibration Procedures 4 ...
Page 118: ...116 ...
Page 119: ...5 Theory of Operation 5 ...
Page 159: ...6 Service 6 ...
Page 177: ...General Disassembly 4 3 2 1 6 Chapter 6 Service Disassembly 175 ...
Page 178: ...Internal DMM Disassembly 3 2 1 Chapter 6 Service Disassembly 176 ...
Page 180: ...Additional Chassis Disassembly 1 2 Chapter 6 Service Disassembly 178 ...
Page 181: ...Plug In Module Disassembly Needle nose Pliers 6 Chapter 6 Service Disassembly 179 ...
Page 182: ...180 ...
Page 183: ...7 Replaceable Parts 7 ...
Page 220: ...Chapter 7 Replaceable Parts Manufacturer s List 218 ...
Page 221: ...8 Schematics 8 ...
Page 272: ......