AC CURRENT Gain Verification Test
Configuration:
AC Current
LF 3 HZ:SLOW (in the
Advanced
menu)
1
Make sure you have read “Test Considerations” on page 71.
2
Select Channel 221, set the AC CURRENT function and the 3 Hz
input filter. With the slow filter selected, each measurement takes
7 seconds to complete. Before executing each test, you must press
to enable reading monitoring on the selected channel (or use the
ROUTe:MON
command from the remote interface).
3
Select each range in the order shown below. Compare measurement
results to the appropriate test limits shown in the table. (Be certain
to allow for appropriate source settling.)
Input
Current Frequency
Range
Quick
check
Error from Nominal
24 hour 90 day 1 year
10 mA
[1]
100 mA
[1]
10 mA
1A
[1]
1 kHz
1 kHz
1 kHz
1 kHz
10 mA
100 mA
1 A
1A
Q
±
14
µ
A
±
600
µ
A
±
1.41 mA
±
1.4 mA
±
14
µ
A
±
600
µ
A
±
1.41 mA
±
1.4 mA
±
14
µ
A
±
600
µ
A
±
1.41 mA
±
1.4 mA
[1] Verify only, no adjustment.
Frequency Gain Verification Test
Configuration:
Frequency
6
1
⁄
2
digits
1
Make sure you have read “Test Considerations” on page 71.
2
Select Channel 210, select the FREQUENCY function and set 6
1
⁄
2
digits.
3
Select each range in the order shown below. Compare measurement
results to the appropriate test limits shown in the table. (Be certain to
allow for appropriate source settling.)
Input
Voltage Frequency
Range
Quick
Check
Error from Nominal
24 hour 90 day 1 year
10 mV
[1]
1 V
100 Hz
100 kHz
100 mV
1 V
Q
±
0.06 Hz
±
6 Hz
±
0 .1 Hz
±
10 Hz
±
0 .1 Hz
±
10 Hz
[1] Verify only, No adjustment. For this test, isolate the calibrator’s output from earth ground.
Q
: Quick performance verification test points.
4
Chapter 4 Calibration Procedures
Internal DMM Verification Tests
79
Summary of Contents for 34970A
Page 1: ...Agilent 34970A Data Acquisition Switch Unit Service Guide ...
Page 34: ...32 ...
Page 35: ...2 Quick Start 2 ...
Page 49: ...3 Front Panel Overview 3 ...
Page 62: ...60 ...
Page 63: ...4 Calibration Procedures 4 ...
Page 118: ...116 ...
Page 119: ...5 Theory of Operation 5 ...
Page 159: ...6 Service 6 ...
Page 177: ...General Disassembly 4 3 2 1 6 Chapter 6 Service Disassembly 175 ...
Page 178: ...Internal DMM Disassembly 3 2 1 Chapter 6 Service Disassembly 176 ...
Page 180: ...Additional Chassis Disassembly 1 2 Chapter 6 Service Disassembly 178 ...
Page 181: ...Plug In Module Disassembly Needle nose Pliers 6 Chapter 6 Service Disassembly 179 ...
Page 182: ...180 ...
Page 183: ...7 Replaceable Parts 7 ...
Page 220: ...Chapter 7 Replaceable Parts Manufacturer s List 218 ...
Page 221: ...8 Schematics 8 ...