Thermocouple Reference Junction Adjustments
These adjustments are plug-in module specific and only affect thermo-
couple measurements. The calibration constants created by these
adjustments are stored in non-volatile memory
on the plug-in module
.
1
Connect a 10 k
Ω
(YSI 44031) thermistor to each of the following channels
(a kit of five thermistors is available as Agilent part number 34308A):
For the 34901AChannels 6 and 17
For the 34902AChannels 6 and 11
For the 34908AChannels 6 and 16
Keep the thermistor leads as short as possible. Locate the thermistor
as near to the input connectors as possible.
2
Install the plug-in module in the mainframe in slot 200. Apply power
and allow a 2 hour warm-up.
3
Set 10 k
Ω
thermistor measurements on Channels 206 and 217
(or 206 and 211). Before executing each test, you must press
to enable reading monitoring on the selected channel (or use the
ROUTe:MON
command from the remote interface).
4
Press
to enter the calibration menu. Press
again to
begin the adjustment procedure.
5
Verify the adjustment (
see page 112
).
Module Reference
34901A
4
Chapter 4 Calibration Procedures
Thermocouple Reference Junction (Optional)
113
Summary of Contents for 34970A
Page 1: ...Agilent 34970A Data Acquisition Switch Unit Service Guide ...
Page 34: ...32 ...
Page 35: ...2 Quick Start 2 ...
Page 49: ...3 Front Panel Overview 3 ...
Page 62: ...60 ...
Page 63: ...4 Calibration Procedures 4 ...
Page 118: ...116 ...
Page 119: ...5 Theory of Operation 5 ...
Page 159: ...6 Service 6 ...
Page 177: ...General Disassembly 4 3 2 1 6 Chapter 6 Service Disassembly 175 ...
Page 178: ...Internal DMM Disassembly 3 2 1 Chapter 6 Service Disassembly 176 ...
Page 180: ...Additional Chassis Disassembly 1 2 Chapter 6 Service Disassembly 178 ...
Page 181: ...Plug In Module Disassembly Needle nose Pliers 6 Chapter 6 Service Disassembly 179 ...
Page 182: ...180 ...
Page 183: ...7 Replaceable Parts 7 ...
Page 220: ...Chapter 7 Replaceable Parts Manufacturer s List 218 ...
Page 221: ...8 Schematics 8 ...
Page 272: ......