Valid Gain Adjustment Input Values
Gain adjustment can be accomplished using the following input values.
Function
Range
Valid Calibration Input
Values
DC VOLTS
OHMS, OHMS 4W
DC CURRENT
AC VOLTS [1]
AC CURRENT
Frequency
100 mV to 100 V
300 V
100
Ω
to 10 M
Ω
10 mA to 1 A
10 mV to 100 V
300 V
1 A
Any
0.9 to 1.1 x Full Scale
250 V to 303 V
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
95 V to 303 V
9 mA to 11 mA
Any Input > 100 mV rms,
1 kHz
–
100 kHz
[1] Valid frequencies are as follows:
1 kHz
±
10% for the 1 kHz calibration,
45 kHz – 100 kHz for the 50 kHz calibration,
and 10 Hz
±
10% for the 10 Hz calibration.
4
Chapter 4 Calibration Procedures
Internal DMM Adjustments
83
Summary of Contents for 34970A
Page 1: ...Agilent 34970A Data Acquisition Switch Unit Service Guide ...
Page 34: ...32 ...
Page 35: ...2 Quick Start 2 ...
Page 49: ...3 Front Panel Overview 3 ...
Page 62: ...60 ...
Page 63: ...4 Calibration Procedures 4 ...
Page 118: ...116 ...
Page 119: ...5 Theory of Operation 5 ...
Page 159: ...6 Service 6 ...
Page 177: ...General Disassembly 4 3 2 1 6 Chapter 6 Service Disassembly 175 ...
Page 178: ...Internal DMM Disassembly 3 2 1 Chapter 6 Service Disassembly 176 ...
Page 180: ...Additional Chassis Disassembly 1 2 Chapter 6 Service Disassembly 178 ...
Page 181: ...Plug In Module Disassembly Needle nose Pliers 6 Chapter 6 Service Disassembly 179 ...
Page 182: ...180 ...
Page 183: ...7 Replaceable Parts 7 ...
Page 220: ...Chapter 7 Replaceable Parts Manufacturer s List 218 ...
Page 221: ...8 Schematics 8 ...
Page 272: ......