(Optional)
34903A Relay Contact Resistance Verification
1
Be sure to read “Plug-in Module Test Considerations” on page 87.
2
Install the 34903A module in slot 100. Close Channels 01 through 20.
Remove the module from the 34970A.
3
Measure the resistance from the
CM
terminal to the
NO
terminal on
each channel.
4
Install the module in slot 100. Open Channel 01 through 20.
Remove the module from the 34970A.
5
Measure the resistance from the
CM
terminal to the
NC
terminal on
each channel.
Note:
In general, a new relay should have a contact resistance of less than
0.2
Ω
. Relays with contact resistance in excess of 1.2
Ω
should be replaced.
Module Reference
4
Chapter 4 Calibration Procedures
Relay Verification
101
Summary of Contents for 34970A
Page 1: ...Agilent 34970A Data Acquisition Switch Unit Service Guide ...
Page 34: ...32 ...
Page 35: ...2 Quick Start 2 ...
Page 49: ...3 Front Panel Overview 3 ...
Page 62: ...60 ...
Page 63: ...4 Calibration Procedures 4 ...
Page 118: ...116 ...
Page 119: ...5 Theory of Operation 5 ...
Page 159: ...6 Service 6 ...
Page 177: ...General Disassembly 4 3 2 1 6 Chapter 6 Service Disassembly 175 ...
Page 178: ...Internal DMM Disassembly 3 2 1 Chapter 6 Service Disassembly 176 ...
Page 180: ...Additional Chassis Disassembly 1 2 Chapter 6 Service Disassembly 178 ...
Page 181: ...Plug In Module Disassembly Needle nose Pliers 6 Chapter 6 Service Disassembly 179 ...
Page 182: ...180 ...
Page 183: ...7 Replaceable Parts 7 ...
Page 220: ...Chapter 7 Replaceable Parts Manufacturer s List 218 ...
Page 221: ...8 Schematics 8 ...
Page 272: ......