Calibration Procedure
The following procedure is the recommended method to complete an
instrument calibration.
1
Read “Test Considerations” (
page 71
).
2
Unsecure the instrument for calibration (
page 51
).
3
Perform the
verification
tests to characterize the instrument
(incoming data).
4
Perform the zero
adjustment
procedures.
5
Perform the gain
adjustment
procedures. Perform the
verification
tests to verify the adjustments (outgoing data).
6
Secure the instrument against calibration.
7
Note the new security code and calibration count in the
instrument’s maintenance records.
Aborting a Calibration in Progress
Sometimes it may be necessary to abort a calibration after the
procedure has already been initiated. You can abort a calibration at
any time by turning off the power. When performing a calibration
from the remote interface, you can abort a calibration by issuing a
remote interface device clear message.
C A U T I O N
If you abort a calibration in progress when the instrument is attempting
to write new calibration constants to EEPROM, you may lose all
calibration constants for the function. Typically, upon re-applying
power, the instrument will report error
705 Cal:Aborted.
You may
also generate errors 740 through 746. If this occurs, you should not use
the instrument until a complete re-adjustment has been performed.
Chapter 4 Calibration Procedures
Calibration Procedure
70
Summary of Contents for 34970A
Page 1: ...Agilent 34970A Data Acquisition Switch Unit Service Guide ...
Page 34: ...32 ...
Page 35: ...2 Quick Start 2 ...
Page 49: ...3 Front Panel Overview 3 ...
Page 62: ...60 ...
Page 63: ...4 Calibration Procedures 4 ...
Page 118: ...116 ...
Page 119: ...5 Theory of Operation 5 ...
Page 159: ...6 Service 6 ...
Page 177: ...General Disassembly 4 3 2 1 6 Chapter 6 Service Disassembly 175 ...
Page 178: ...Internal DMM Disassembly 3 2 1 Chapter 6 Service Disassembly 176 ...
Page 180: ...Additional Chassis Disassembly 1 2 Chapter 6 Service Disassembly 178 ...
Page 181: ...Plug In Module Disassembly Needle nose Pliers 6 Chapter 6 Service Disassembly 179 ...
Page 182: ...180 ...
Page 183: ...7 Replaceable Parts 7 ...
Page 220: ...Chapter 7 Replaceable Parts Manufacturer s List 218 ...
Page 221: ...8 Schematics 8 ...
Page 272: ......