34907A Analog Output
Analog Output Verification Test
This procedure is used to check the calibration of the analog outputs
on the 34907A Multifunction Module. Install the module in slot 200.
Verification checks are performed only for those output values with
unique calibration constants.
1
Make connections to analog output channels as shown below.
2
For each analog output, set each output value in the table below.
Compare measurement results to the appropriate test limits shown
in the table.
DAC
Output
Quick
Check
Measured
Output
Error from Nominal
1 year
0.000 V
10.000 V
Q
0.000 V
10.000 V
±
20 mV
±
45 mV
34907A
To
DMM
To DMM
Chapter 4 Calibration Procedures
34907A Analog Output
114