To Verify the Battery
1
Remove AC line power (this also provides a load on A1BT101).
2
Remove the cover (
see page 174
).
3
Measure the battery voltage as shown. Replace the battery if
the voltage is below 2.7 V.
To Replace the Battery
1
Remove AC line power.
2
Remove the cover (see page 174). If installed, remove the
Internal DMM (A4) assembly (
see page 176
).
3
Turn the instrument over and unsolder the three battery terminals
from the bottom of the circuit board. Use proper through-hole
soldering techniques and equipment. Remove the battery and
dispose of properly.
4
Install, solder, and verify the new battery.
Battery –
B
6
Chapter 6 Service
Battery Check and Replacement
173