Agilent Technologies Calibration Services
When your instrument is due for calibration, contact your local
Agilent Service Center for a low-cost recalibration. The 34970A is
supported on automated calibration systems which allow Agilent to
provide this service at competitive prices.
Calibration Interval
The instrument should be calibrated on a regular interval
determined by the measurement accuracy requirements of your
application.
A 1-year interval is adequate for most applications. Accuracy
specifications are warranted
only
if adjustment is made at regular
calibration intervals. Accuracy specifications are not warranted
beyond the 1-year calibration interval. Agilent does not recommend
extending calibration intervals beyond 2 years for any application.
Adjustment is Recommended
Whatever calibration interval you select, Agilent recommends that
complete re-adjustment should always be performed
at the calibration
interval. This will assure that the 34970A will remain within
specification for the next calibration interval. This criteria for
re-adjustment provides the best long-term stability. Performance
data measured using this method
can
be used to extend future
calibration intervals.
Use the Calibration Count feature (
see page 69
) to verify that all
adjustments have been performed.
4
Chapter 4 Calibration Procedures
Agilent Technologies Calibration Services
63
Summary of Contents for 34970A
Page 1: ...Agilent 34970A Data Acquisition Switch Unit Service Guide ...
Page 34: ...32 ...
Page 35: ...2 Quick Start 2 ...
Page 49: ...3 Front Panel Overview 3 ...
Page 62: ...60 ...
Page 63: ...4 Calibration Procedures 4 ...
Page 118: ...116 ...
Page 119: ...5 Theory of Operation 5 ...
Page 159: ...6 Service 6 ...
Page 177: ...General Disassembly 4 3 2 1 6 Chapter 6 Service Disassembly 175 ...
Page 178: ...Internal DMM Disassembly 3 2 1 Chapter 6 Service Disassembly 176 ...
Page 180: ...Additional Chassis Disassembly 1 2 Chapter 6 Service Disassembly 178 ...
Page 181: ...Plug In Module Disassembly Needle nose Pliers 6 Chapter 6 Service Disassembly 179 ...
Page 182: ...180 ...
Page 183: ...7 Replaceable Parts 7 ...
Page 220: ...Chapter 7 Replaceable Parts Manufacturer s List 218 ...
Page 221: ...8 Schematics 8 ...
Page 272: ......