Each analog-to-digital conversion begins when the multimeter is
triggered. The ADC starts by clearing the integrator slope count in
A1U209. At the end of the integration period, the slope count is
latched. The slope count provides the most significant bits of the
input voltage conversion. The least significant bits are converted by
the on-chip ADC of CPU A1U205.
The instrument precision voltage reference is U403. Resistor R409
provides a stable bias current for the reference zener diode. R408
and CR404 provide a bias to assure that the reference zener biases to
+7 V during power up. IC U400A amplifies the voltage reference to
+10 V while amplifier U401A inverts the +10 V reference to –10 V.
The reference voltages force precision slope currents for the
integrating ADC through U102E–R17, R18.
5
Chapter 5 Theory of Operation
Internal DMM
137
Summary of Contents for 34970A
Page 1: ...Agilent 34970A Data Acquisition Switch Unit Service Guide ...
Page 34: ...32 ...
Page 35: ...2 Quick Start 2 ...
Page 49: ...3 Front Panel Overview 3 ...
Page 62: ...60 ...
Page 63: ...4 Calibration Procedures 4 ...
Page 118: ...116 ...
Page 119: ...5 Theory of Operation 5 ...
Page 159: ...6 Service 6 ...
Page 177: ...General Disassembly 4 3 2 1 6 Chapter 6 Service Disassembly 175 ...
Page 178: ...Internal DMM Disassembly 3 2 1 Chapter 6 Service Disassembly 176 ...
Page 180: ...Additional Chassis Disassembly 1 2 Chapter 6 Service Disassembly 178 ...
Page 181: ...Plug In Module Disassembly Needle nose Pliers 6 Chapter 6 Service Disassembly 179 ...
Page 182: ...180 ...
Page 183: ...7 Replaceable Parts 7 ...
Page 220: ...Chapter 7 Replaceable Parts Manufacturer s List 218 ...
Page 221: ...8 Schematics 8 ...
Page 272: ......