Internal DMM Verification Tests
These procedures use inputs connected to a 34901A 20-Channel
Multiplexer (
see page 66
) installed in slot 200.
Zero Offset Verification
This procedure is used to check the zero offset performance of the
internal DMM. Verification checks are only performed for those
functions and ranges with unique offset calibration constants.
Measurements are checked for each function and range as described
in the procedure below.
Zero Offset Verification Procedure
1
Make sure you have read “Test Considerations” on page 71.
2
This procedure will measure the shorts installed on Channels 209
and 219. Leave the Amps input connections (Channel 221) open.
Continued on next page...
4
Chapter 4 Calibration Procedures
Internal DMM Verification Tests
75
Summary of Contents for 34970A
Page 1: ...Agilent 34970A Data Acquisition Switch Unit Service Guide ...
Page 34: ...32 ...
Page 35: ...2 Quick Start 2 ...
Page 49: ...3 Front Panel Overview 3 ...
Page 62: ...60 ...
Page 63: ...4 Calibration Procedures 4 ...
Page 118: ...116 ...
Page 119: ...5 Theory of Operation 5 ...
Page 159: ...6 Service 6 ...
Page 177: ...General Disassembly 4 3 2 1 6 Chapter 6 Service Disassembly 175 ...
Page 178: ...Internal DMM Disassembly 3 2 1 Chapter 6 Service Disassembly 176 ...
Page 180: ...Additional Chassis Disassembly 1 2 Chapter 6 Service Disassembly 178 ...
Page 181: ...Plug In Module Disassembly Needle nose Pliers 6 Chapter 6 Service Disassembly 179 ...
Page 182: ...180 ...
Page 183: ...7 Replaceable Parts 7 ...
Page 220: ...Chapter 7 Replaceable Parts Manufacturer s List 218 ...
Page 221: ...8 Schematics 8 ...
Page 272: ......