Gain Adjustment Procedure
Adjustment for each function should be performed
only
in the order
shown in the performance verification table. The performance
verification tables used for gain adjustments start on page 77.
Review the “Test Considerations” (page 71) and “Gain Adjustment
Considerations” (page 82) sections before beginning this test.
Configuration:
DC functions — 6
1
⁄
2
digits
AC functions — LF 3 HZ:SLOW (in the
Advanced
menu)
1
Select Channel 210. Configure the channel to each function and
range shown in the gain verification tables (
pages 75 – 79
).
2
Apply the input signal shown in the “Input” column of the
appropriate verification table.
Note:
Always complete tests in the same order as shown in the
appropriate verification table.
3
Press
to enter the calibration menu. Press
again to
begin the adjustment procedure.
4
Use the knob,
and
to set the number in the display to the
actual input value and press
.
5
Perform the appropriate Gain Verification Test to check the
calibration results.
6
Repeat steps 1 through 6 for each gain verification test point shown
in the tables.
Note:
Each range in the gain adjustment procedure takes less than
20 seconds to complete.
Chapter 4 Calibration Procedures
Internal DMM Adjustments
84
Summary of Contents for 34970A
Page 1: ...Agilent 34970A Data Acquisition Switch Unit Service Guide ...
Page 34: ...32 ...
Page 35: ...2 Quick Start 2 ...
Page 49: ...3 Front Panel Overview 3 ...
Page 62: ...60 ...
Page 63: ...4 Calibration Procedures 4 ...
Page 118: ...116 ...
Page 119: ...5 Theory of Operation 5 ...
Page 159: ...6 Service 6 ...
Page 177: ...General Disassembly 4 3 2 1 6 Chapter 6 Service Disassembly 175 ...
Page 178: ...Internal DMM Disassembly 3 2 1 Chapter 6 Service Disassembly 176 ...
Page 180: ...Additional Chassis Disassembly 1 2 Chapter 6 Service Disassembly 178 ...
Page 181: ...Plug In Module Disassembly Needle nose Pliers 6 Chapter 6 Service Disassembly 179 ...
Page 182: ...180 ...
Page 183: ...7 Replaceable Parts 7 ...
Page 220: ...Chapter 7 Replaceable Parts Manufacturer s List 218 ...
Page 221: ...8 Schematics 8 ...
Page 272: ......