Performance Verification Tests
Use the Performance verification Tests to verify the measurement
performance of the instrument. The performance verification tests
use the instrument’s specifications listed in chapter 1, “Specifications,”
starting on page 15.
You can perform four different levels of performance verification tests:
•
Self-Test
A series of internal verification tests that give a high
confidence that the instrument is operational.
•
Quick Verification
A combination of the internal self-tests and
selected verification tests.
•
Performance Verification Tests
An extensive set of tests that
are recommended as an acceptance test when you first receive the
instrument or after performing adjustments.
•
Optional Verification Tests
Tests
not
performed with every
calibration. Perform these tests to verify additional specifications
or functions of the instrument.
Chapter 4 Calibration Procedures
Performance Verification Tests
72
Summary of Contents for 34970A
Page 1: ...Agilent 34970A Data Acquisition Switch Unit Service Guide ...
Page 34: ...32 ...
Page 35: ...2 Quick Start 2 ...
Page 49: ...3 Front Panel Overview 3 ...
Page 62: ...60 ...
Page 63: ...4 Calibration Procedures 4 ...
Page 118: ...116 ...
Page 119: ...5 Theory of Operation 5 ...
Page 159: ...6 Service 6 ...
Page 177: ...General Disassembly 4 3 2 1 6 Chapter 6 Service Disassembly 175 ...
Page 178: ...Internal DMM Disassembly 3 2 1 Chapter 6 Service Disassembly 176 ...
Page 180: ...Additional Chassis Disassembly 1 2 Chapter 6 Service Disassembly 178 ...
Page 181: ...Plug In Module Disassembly Needle nose Pliers 6 Chapter 6 Service Disassembly 179 ...
Page 182: ...180 ...
Page 183: ...7 Replaceable Parts 7 ...
Page 220: ...Chapter 7 Replaceable Parts Manufacturer s List 218 ...
Page 221: ...8 Schematics 8 ...
Page 272: ......