34908A
Components in this discussion are located on the A1 circuit assembly
(34908-66501).
The schematics are included in Chapter 8 starting on
page 267.
The control circuitry has three grouping of latches, relay drivers and
relays. The 40 voltage and resistance measurement channels are, for
control purposes, arranged into 8 rows by 10 columns. Tree switching
controls bank selection and connections to the analog bus.
The row latch, U102, and column latches, U103 and U104, control the
relays. The row drivers are divided into four groups of set and reset
drivers. Each group of row drivers controls five relays. The column
drivers operate as a pair. There are ten column drivers each
controlling two relays. The analog bus backplane relays are
non-latching.
Module Reference
5
Chapter 5 Theory of Operation
Switch Modules
149
Summary of Contents for 34970A
Page 1: ...Agilent 34970A Data Acquisition Switch Unit Service Guide ...
Page 34: ...32 ...
Page 35: ...2 Quick Start 2 ...
Page 49: ...3 Front Panel Overview 3 ...
Page 62: ...60 ...
Page 63: ...4 Calibration Procedures 4 ...
Page 118: ...116 ...
Page 119: ...5 Theory of Operation 5 ...
Page 159: ...6 Service 6 ...
Page 177: ...General Disassembly 4 3 2 1 6 Chapter 6 Service Disassembly 175 ...
Page 178: ...Internal DMM Disassembly 3 2 1 Chapter 6 Service Disassembly 176 ...
Page 180: ...Additional Chassis Disassembly 1 2 Chapter 6 Service Disassembly 178 ...
Page 181: ...Plug In Module Disassembly Needle nose Pliers 6 Chapter 6 Service Disassembly 179 ...
Page 182: ...180 ...
Page 183: ...7 Replaceable Parts 7 ...
Page 220: ...Chapter 7 Replaceable Parts Manufacturer s List 218 ...
Page 221: ...8 Schematics 8 ...
Page 272: ......