A 600
Options
GB
02446
-
E
d
it
io
n
04
-
J
une
08
Metal seals
1
Inlet and high vacuum manifolds and the analyzer cell are equipped
with metal seals instead of elastomer seals to protect the leak
detector against contamination with helium. This option is particularly
usefull in case of high sensitivity helium leak detection in an ”helium
contaminated environment”.
Localisation of the metal seals
F 800
Inlet port
2
ASM 122 D: The standard DN 25 inlet port can be replaced by a
DN 40 inlet port for convenience.
ASM 1002: The test chamber can be replaced by a DN 25 inlet
port for convenience.
Units
3
The user can choice the unit of the software: mbar.l/s, Pa.m
3
/s or
Torr.l/s.
Languages
4
The user can choice the language of the software: English, French,
German or Japanese.
Note: ASM 142 S: English/French/German/Spanish.
ASM 1002: English/French.
3 masses
5
For use of one of the 3 following tracer gases:
Helium 4, Helium 3 or Hydrogen 2.
Automatic test
chambers
6
This is used for the automatic bombing testing of small components.
When the chamber cover is closed, the test cycle is initiated, via a
contact.
3 aluminium alloy models are available:
•
a hemispheric chamber, Ø 72 mm, depth 31 mm (small model),
•
a cylindrical chamber, maximum Ø 85 mm and
maximum depth 68 mm (medium model),
•
a cylindrical chamber, maximum Ø 160 mm and
maximum depth 200 mm (large model).
Note: ASM 142: large model not available.