E 111
Access to the internal components
ASM 142 D - ASM Graph D+
G
B
01490
-
E
d
it
io
n
03
-
J
un
e
08
Access level 2 (ctd)
Disassemble the main block
Remove all the connections:
2 connections for analyzer cell
1
.
2 quick connectors (for valves)
2
.
1
2
Molecular pump connector
3
.
Exhaust and intermediate port connections on molecular pump
4
.
Primary pump connectors
5
.
3
5
(ASM 142 D)
4
(ASM 142 D)
3
5
4
(ASM Graph D+)
(ASM 142 D)
3/5