C 302
Advanced internal calibration of the leak
detector
1/3
GB
00970
-
E
d
it
io
n
05
-
J
une
08
A help for control panel utilization/access.
Operating principle of the control panel
C 110
Access to level
4
- Password
C 130
Setting and maintenance part presentation of the control panel
Access to parameters and parameters active
depending on authorization
C 120
Summary of screens
Complete displays list with access
way and associated sheet
C 140
Introduction
The operator has the possibility to control on the internal calibration
process:
•
activation / deactivation of the internal calibration.
•
setting of the checking function.
Activation/
deactivation of the
internal calibration
Procedure
For maintenance for example.
Adjustment
C 110
Notes:
•
If the internal calibration is deactivated (off), it is still possible to
start a test cycle and use the leak detector.
•
When internal calibration function is activated (on), user could
activate or not activate the checking function.