C 200
CAL IN PROGRESS
ELECTR.ZERO ADJUST.
80
READY FOR CYCLE
...
...
...
GB
02460
-
E
d
it
io
n
02
-
J
une
08
Starting up / Switching off the leak
detector
Starting up of the leak
detector
(continued)
When the pumps have reached their nominal speed, the unit auto-
calibrates: different screens will show.
1
When calibration is completed,
the unit is ready to start a cycle.
...
STDBY
The digital voice gives to the operator the message:
“detector ready for cycle”.
Flashing
When the filament is OK, the unit
is ready to start a cycle.
A « i » is flashing at the right end
of the 1st line of the LCD.
READY FOR CYCLE
...
...
i
...
It informs the operator that the
unit is in manual calibration:
2
it will remain in manual
calibration until operator
performs an
auto-calibration.
STDBY
Information message
C 580
Internal calibration on request
C 300
Note: It is possible to start a test cycle even if the detector is not
calibrated.
Switching off
the leak detector
The leak detector can be switched off at any time by depressing the
main power switch to 0 (OFF) but it is recommended (especially when
the leak detector may not be used for an extended period of time) to
stop it following the procedure described below.