Replacement of the internal calibrated leak
GB
01253
-
E
d
it
io
n
04
-
J
une
08
Installation of
a new internal
calibrated leak
After recalibration, your «recalibrated» leak is returned to you with:
- its new value,
- a calibration certificate.
To install the calibrated internal leak (new or recalibrated):
Switch off the detector and unplug it from the main power.
Install the leak:
- connect the flange,
- connect the heat sensor.
Entering the new internal calibrated leak values.
Calibrated leak values programming
C 305
Close the leak detector.
Connect the detector to the main power.
Switch it on: a calibration is performed automatically at the end of
the start-up sequence if the autocal is on.
It is recommended to repeat a calibration after 1 hour of operation
when the temperature has stabilized inside the unit.
Basic internal calibration of the leak detector
C 301
Intensive use of the
detector
In the case of intensive use of the detector, it is recommended to have
a spare internal calibrated leak.
If this is not possible, the detector can still be used and calibrated
using an external calibrated leak.
External calibration of the leak detector
C 303