The control panel can be divided into two different sections.
•
The section located on the right of the control panel is dedicated to
the user.
•
The section located on the left of the control panel dedicated to the
setting and maintenance (adjustments, functions, menu access, etc.).
Operation part
F1
F2
LEAK RATE
F3
COR
mbar.l/s
Pa.m
3
/s
Torr.l/s
F4
10
-12
10
-11
10
-10
10
-9
10
-8
10
-7
10
-6
10
-5
10
-4
10
-3
10
-2
10
-3
10
-2
10
-1
1 10
1
10
2
10
3
mbar
hPa
Torr
STDBY
EVAC TEST
NEXT
-
+
RESET
INLET PRESSURE
SET POINTS
SPECTRO
MAINTENANCE
OTHER
AUTO CAL
SNIFFER
ZERO
CYCLE
Setting and maintenance part
Levels
Description
•
The detector offers 4 user interface levels for this section to
accomodate any application requirements.
Setting and maintenance part
User part
Level
1
This level has very limited information on the
alphanumeric display (LCD). This level is generally
selected for production types of applications.
No access to control keys
(Cycle key included)
Level
2
This level allows the operator to visualize some
parameters without the possibility of making any
changes.
Same as Level
1
, this level is usually selected for
production types of applications.
Access to all the control
keys
Level
3
Same as level
2
but with possibility to set some
parameters.
This level is generally selected for maintenance
applications.
Level
4
This level allows access to all the parameters and is
generally used for settings all the parameters.
Note: When switching from level
4
to any other
level, the switch can be performed without using the
password.
This level is generally selected for R&D applications.