C 301
Basic internal calibration of the leak
detector
4/5
PLEASE,OPEN He LEAK
LEAK VALUE: 1.0E-07
EXT
θ
(°C/°F):20/68
OK -> press NEXT
GB
00969
-
E
d
it
io
n
06
-
J
une
08
3
Start the
Place the external calibrated
calibration
leak on the test port.
Start the calibration.
4
Calibration
preparation
These 2 screens appear alternately.
ON TEST PORT!
Leak value: 1.0E-07
EXT
θ
(°C/°F):20/68
OK -> press NEXT
Note: If the operator press
,all the
parameter of the programmed calibrated leak are
displayed.
C 305
Open the valve of the calibrated leak (if there is one).
Set the ambiant temperature.
Adjustment
C 110
5
Validate the
calibration
process
NEXT
Different screens appear during the process giving
internal parameters values.