C 302
Advanced internal calibration of the leak
detector
3/3
GB
00970
-
E
d
it
io
n
05
-
J
une
08
Checking function with an
external calibrated leak
It is possible to perform a calibration checking with an external
calibrated leak (position parameter = external). But it is not automatic
and in the case, the leak detector performs a complete calibration (not
only a comparison like with the internal calibrated leak).
When the first of the 2 counters is reached, a “i“ will flash at the right
end of the 1st line of the LCD.
READY FOR CYCLE
i
...
...
...
The “i” flashing will stay on the LCD until operator starts an external
calibration.
C 303
An audio message advises the operator in the process to follow.
Note: It is still possible to start a test cycle even without performing an
internal calibration .
On request calibration
checking
C 301