Set up your insulin pump with a new cartridge, adapter and
infusion set and prime.
Please refer to section 4 “Prepare the cartridge and attach the
adapter and infusion set” for further information. When
finished, put your insulin pump into
RUN
(press
d
and
f
).
5.
If the adapter is connected incorrectly (with inserted car-
tridge):
Remove the adapter and infusion set from your in-
sulin pump.
Set up your insulin pump with a new adapter and infusion set
and prime.
Please refer to section 4 “Prepare the cartridge and attach the
adapter and infusion set” for further information. When
finished, put your insulin pump into
RUN
(press
d
and
f
).
6.
If the piston rod is accidentally returned or not in contact
with the cartridge plunger (with correctly connected
adapter):
Remove the infusion set from your insulin pump.
Set up your insulin pump with a new infusion set and prime.
Please refer to section 4 “Prepare the cartridge and attach the
adapter and infusion set” for further information. When
finished, put your insulin pump into
RUN
(press
d
and
f
).
7.
If the cartridge is leaking (with piston rod in position):
Remove the cartridge, adapter and infusion set from your in-
sulin pump.
If the cartridge is broken or leaking, clean and dry the car-
tridge compartment of your insulin pump.
Please refer to section 10.2.2 “Your insulin pump” for further
information.
Set up your insulin pump with a new cartridge, adapter and
infusion set and prime.
Caution:
Broken cartridges can cause injury (glass
splinters).
Info:
Insulin is recognizable by its unique smell
due to its preservatives, therefore, a defective car-
tridge may be recognized by smelling for leaking
insulin in the cartridge compartment.
8.
If the adapter is leaking (with undamaged cartridge):
Remove the adapter and infusion set tubing from your insulin
pump.
Set up your insulin pump with a new adapter and infusion set
tubing and prime.
Please refer to section 4 “Prepare the cartridge and attach the
adapter and infusion set” for further information.
When finished, reconnect the infusion set tubing and put your
insulin pump into
RUN
(press
d
and
f
).
If the “Alert A4: Cartridge/Adapter Alert”occurs again, repeat the process using only
new disposables (cartridge, adapter, infusion set). If the “Alert A4: Cartridge/Adapter
Alert” occurs again despite the use of new components, please contact your
Accu-Chek customer service center for insulin pumps.
With a non-disconnectable infusion set do the following:
1. Silence and confirm the alert by pressing
f
twice.
2.
Check your blood glucose level. If it is high, take the
appropriate actions according to your health care
team’s instructions.
3. Remove the infusion set from the site.
4.
If no cartridge is in the cartridge compartment:
Remove the
adapter and infusion set from your insulin pump.