Accu-Chek D-TRONplus insulin pump Reference Manual
145
Accu-Chek D-TRONplus insulin pump Reference Manual
144
No.
Meaning of the error
What you have to do
Page
E1
Empty Cartridge
Silence and confirm the error.
146
Act according to section 7.2.1
“Error E1: Empty Cartridge”.
When finished, put your insulin pump
into
RUN
.
E2
PowerPack Depleted
Silence and confirm the error.
147
Act according to section 7.2.2
“Error E2: PowerPack Depleted”.
When finished, put your insulin pump
into
RUN
.
E3
Automatic Off
Silence and confirm the error.
147
Put your insulin pump into
RUN
.
E4
Occlusion
Silence and confirm the error.
148
Act according to section 7.2.4
“Error E4: Occlusion”.
When finished, put your insulin pump
into
RUN
.
E5
End of Use Time
Silence and confirm the error.
150
(t1 or
Act according to section 7.2.5
t2)
“Error E5: End of Use Time”
E6
Mechanical Error
Silence and confirm the error.
150
Act according to section 7.2.6
“Error E6: Mechanical Error”.
E7
Electronic Error
Act according to section 7.2.7
151
“Error E7: Electronic Error”.
7.2 Overview of the errors
7.1.7 Alert A7: Temporary Basal Rate Over
An “Alert A7: Temporary Basal Rate Over” occurs when a temporary basal rate has
ended. Once the alert has occured, the original basal rate (100%) resumes automa-
tically.
Do the following:
1. Silence and confirm the alert by pressing
f
twice.
2. Decide whether a further temporary basal rate change is
appropriate and program one if necessary.
If desired, the “Alert A7: Temporary Basal Rate Over” can be deactivated using the
Accu-Chek DiaLog Pump Programming tool. If deactivated, the “Alert A7: Temporary
Basal Rate Over” does not occur when the temporary basal rate ends.
7.1.8 Alert A8: Bolus Cancelled
If a bolus is cancelled after the bolus delivery has begun, the bolus delivery is can-
celled and an “Alert A8: Bolus Cancelled” occurs.
Please refer to section 5.2 “Program a bolus” for further information.
Do the following:
1. Silence and confirm the alert by pressing
f
twice.
2. Put your insulin pump into
RUN
(press
d
and
f
) if
necessary.
3. Ensure that the cancellation was intended and program
a new bolus if necessary.
The bolus amount delivered can be reviewed in the bolus history.
Please refer to section 6.1.1 “Review the bolus history” for further information.