2.5 Standard configuration
Warning: Incorrect programming of your insulin pump
may
cause inappropriate insulin delivery. Your insulin pump must be pro-
grammed with your personal settings prior to starting insulin pump the-
rapy. Do not operate your insulin pump without knowing your personal
settings. If you are not sure about your personal settings or you are less
experienced, have your doctor and/or your health care team check your
personal settings. Work with your doctor and/or your health care team to
determine the timing, amount and type of bolus you need to deliver. Make
sure you know your personal insulin/carbohydrate ratio and your correc-
tion bolus ratio.
Warning:
Contact your doctor and/or your health care team for an alter-
nate therapy plan when you interrupt the insulin pump therapy for a
longer period of time.
Warning:
Before using another insulin pump, always check your perso-
nal settings before use in order to avoid incorrect insulin delivery. Please
record your personal settings from your current pump and ensure that
they are correctly programmed into a new pump. Make sure that the time
and date are set correctly. If you are not sure about your personal settings
or you are less experienced, contact your doctor and/or your health care
team to check your personal settings.
Your insulin pump is provided with standard configuration (factory settings). You may
need to adapt these parameters to meet your individual needs. Always consult your
doctor or health care team before adapting parameters.
All changes which can be made on your insulin pump can also be made on a personal
computer (PC) using Accu-Chek DiaLog Pump Programming tool (except KeyLock).
Overview of Accu-Chek D-TRONplus insulin pump functions
The graphic below provides an overview of the functions and their arrangement:
Info:
The functions for daily use are accessible from the
RUN
operating
status, while functions such as returning the piston rod, priming the infu-
sion set and data transfer are accessible from the
STOP
operating status.
KeyLock activated and keys locked
• standard bolus
• remaining cartridge
content
• backlight
• scroll bolus
• extended bolus
• review or set the time
and date
• automatic off
• adjust the beep volume
• adjust the bolus
increment
• lock or unlock the basal
rate programming and
profile selection
• activate or deactivate
the KeyLock
• review the remaining
time (t1)
• review the remaining
time (t2)
• end
• remaining
cartridge
content
• backlight
data transfer
operating status RUN
bolus menu / history
stop insulin delivery
select basal rate profile
program basal rate profile A
program basal rate profile B
program a temporary basal rate
history of daily totals
history of alerts and errors
SETUP menu
operating status STOP
start insulin delivery
return piston rod
prime infusion set