100
Sensitivity adjustment setting
Display settings: MENU key
Sensitivity adjustment button
Settings button
[Sensitivity adjustment setting] screen
1
External weight value input
2
Enter the correct internal weight value button
Name
Setting value
(setting range)
Description
1
External weight value input
Input range
*1
Input the external weight value. This is
common to the “Adjustment weight value”
input item in the [External sensitivity
adjustment] screen.
2
Enter the correct internal weight value
button
-
Displays the [Internal sensitivity
adjustment] screen.
Settings in the red box are default values (factory settings).
*1
Input range
BA-6TE/BA-6DTE
: 0.9 g ~
5 g ~
5.1 g
BA-225TE/BA-225DTE
: 9.9 g ~ 200 g ~ 200.1 g
BA-125DTE
: 9.9 g ~ 100 g ~ 100.1 g
Correcting the internal weight value
Display settings: MENU key
Sensitivity adjustment button
Settings button
Enter the correct internal weight value button [Internal sensitivity adjustment] screen
1
Back button
Name
Description
1
Back button
Returns to the previous screen.
This function corrects the internal weight value based on your external weight. Execute the external
sensitivity adjustment in advance.
The balance automatically loads and unloads the internal weight and corrects the internal weight value.
Then the balance automatically executes the internal sensitivity adjustment. When the adjustment is
complete, the [Sensitivity adjustment setting] screen is displayed.
The corrected value is stored in nonvolatile memory even if the AC adapter is removed.
Summary of Contents for BA-T Series
Page 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Page 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Page 117: ...117 Output example for periodic check output results Periodic check output results...
Page 118: ...118 Periodic check results PDF...
Page 129: ...129 AND MEET analysis graph example...
Page 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...
Page 219: ......