16
Precautions during use for more accurate weighing
For precise and accurate weighing, please take notice of the
following.
Weighing errors may occur due to the influence of static
electricity. Note that if the ambient humidity drops below 45% RH
insulators such as plastics are liable to have static electricity.
Ground the balance and perform the following as needed.
A DC type ionizer that produces no currents of air is available for
the BA-T series balances. Refer to “
” and remove
static electricity from the charged sample directly.
Increase the relative humidity at the place where the
balance is installed.
Weigh the sample in a conductive metal container or the
like.
Wipe off charged materials such as plastic with a damp
cloth to suppress static electricity.
Influence of magnetism may cause weighing errors.
When measuring magnetic materials (iron, etc.), keep the
sample away from the balance main body by means such as
underhook weighing.
Place the sample in the center of the weighing pan.
Weighing errors may occur if there is a difference between the
ambient temperature and temperature of the sample (and the
container). For example, when the room temperature is 20 °C,
convection occurs around a Petri dish or watch glass that is
40 °C and the balance displays a value lighter than the actual
weight. Before weighing the sample and the container, try to
acclimatize them to the ambient temperature.
Perform the weighing operation carefully and quickly. If
measurement takes a long time, error-inducing factors will
increase due to changes in temperature and humidity in the
weighing chamber, air turbulence or reaction/humidity absorption
by the sample.
Do not leave the sample on the weighing pan for an extended
period of time. If a sample is left on the weighing pan for a long
time, the measured value will change due to deviation from the
zero point caused by environmental changes or due to creep
phenomenon.
Metal case
Charged sample
Ground
Ionizer
Magnetic material
Underhook
weighing
Room
temperature 20 °C
Convection
40 °C
No impact shock
Good
Bad
When placing a sample on the weighing pan, do not drop it, or do not place a sample greater than the
balance weighing capacity. In addition, place the sample in the center of the pan.
With the highly sensitive BA-6TE, the weighing value may fluctuate after the stabilization indicator lights
up. It is advisable to set a certain reading time (10 seconds, etc.) after the stabilization indicator lights
up and read the value.
Summary of Contents for BA-T Series
Page 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Page 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Page 117: ...117 Output example for periodic check output results Periodic check output results...
Page 118: ...118 Periodic check results PDF...
Page 129: ...129 AND MEET analysis graph example...
Page 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...
Page 219: ......