11
Assembly and installation
The BA-6TE is used here for the example of assembly. Other models can be assembled in the same
way.
Step 1. Pull out the breeze break
locking handles.
Step 2. Place the breeze break unit on
the weighing unit.
Step 3. Push in the breeze break locking
handles to secure the units to each
other.
Step 4. Assemble the weighing pan.
For BA-6TE / BA-6DTE, assemble
the five parts of the weighing pan in
the correct position by referring to
the cross-sectional view.
In addition to the standard weighing pan (Pan
A), you can also use the weighing
pan for filters. The weighing pan for
filters is approximately 0.2 g heavier
than the standard weighing pan
(Pan A). Therefore, the weighing
capacity when the weighing pan for
filters is used will be approximately
6.0 g.
For BA-225TE / BA-225DTE /
BA-125DTE, assemble the four
parts of the weighing pan in the
correct position by referring to the
cross-sectional view.
Breeze break unit
Step 1
.
Step 3
.
Step 2
.
Weighing unit
For BA-225TE / BA-225DTE / BA-125DTE
For BA-6TE / BA-6DTE
Cross-sectional views
BA-6TE / BA-6DTE
Pan A
Pan support A
Breeze
break ring A
Pan B
Breeze break
ring B
Dust plate A
Weighing pan
for filters
Dust plate B
Breeze break
bottom plate
Cross-sectional view
BA-225TE / BA-225DTE / BA-125DTE
Weighing pan
for filters
Pan A
Summary of Contents for BA-T Series
Page 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Page 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Page 117: ...117 Output example for periodic check output results Periodic check output results...
Page 118: ...118 Periodic check results PDF...
Page 129: ...129 AND MEET analysis graph example...
Page 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...
Page 219: ......