15
Bubble
spirit level
The balance is
level (Correct
weighing)
The balance is
tilted (Inaccurate
weighing)
Up
Leveling foot
Down
Down
Leveling foot
Up
Red circle
Bubble
Be sure to warm up the balance before use for at least an hour, or at least four hours for BA-6TE/BA-
6DTE, with the AC adapter connected to the power supply.
Adjust the sensitivity of the balance before using it for the first time or after having moved it to another
location so that accurate weighing can be performed. For how to adjust sensitivity, refer to
“
10. [Sensitivity Adjustment] Screen
Do not install the balance where flammable or corrosive gas is present.
How to adjust the level of the balance
Level the balance by adjusting the leveling feet
so that the bubble of the bubble spirit level is
centered in the red circle.
Leveling foot
Bubble spirit level
When the bubble is off to the left.
Turn the leveling foot on the front right in the
clockwise direction.
When the bubble is off to the right.
Turn the leveling foot on the front left in the
clockwise direction.
When the bubble is off to the
backward position.
Turn both leveling feet on the front in the clockwise
direction at the same time.
When the bubble is off to the
forward position.
Turn both leveling feet on the front in the
counterclockwise direction at the same time.
Summary of Contents for BA-T Series
Page 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Page 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Page 117: ...117 Output example for periodic check output results Periodic check output results...
Page 118: ...118 Periodic check results PDF...
Page 129: ...129 AND MEET analysis graph example...
Page 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...
Page 219: ......