84
8. Password Function
Intended use
The password function allows you to restrict the use and functions of the balance.
It is effective in preventing falsification of date and time settings or preventing changes in the function
table by the user.
Operation
To log in, press the LOG-IN key
when the display is on, input the user name and password in the
[Log-in] screen, and press the Log-in button
.
To log out, press the Log-out button
or ON:OFF key
to turn the display off.
How to manage balance functions/usage and factory settings
User level
You can restrict functions at each user level.
At the factory setting, all functions are allowed at each user level.
There are four user levels: Administrator, Lab manager, Supervisor and Operator.
Users can select the Lab manager and Supervisor user levels.
Users who are not logged in are an Operator. (Users who do not use the password function)
Administrator
Only the Administrator can restrict functions. Refer to “
The Administrator is registered before shipment and you cannot delete this user or change the user
level of this user.
The password of the Administrator is “0000” at the factory setting. Be sure to change the password of
the Administrator if you want to use the password function. If the password of the Administrator is lost or
forgotten, you cannot manage users or change user authorization. Contact your local A&D dealer to
reset the password.
Summary of Contents for BA-T Series
Page 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Page 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Page 117: ...117 Output example for periodic check output results Periodic check output results...
Page 118: ...118 Periodic check results PDF...
Page 129: ...129 AND MEET analysis graph example...
Page 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...
Page 219: ......