196
AK code and error codes
When [AK (acknowledge), error code] is set to ON in the [Command settings] screen described in
“
”, the balance always responds to reception of all commands sent from a
personal computer or a PLC. Checking the code that is responded improves the reliability of the
communication.
Balance response
When [AK (acknowledge), error code] is set to ON, the balance responds as follows.
When the balance receives a command requesting data:
If the balance can output the data, it sends the requested data.
If the balance cannot output the data, it sends an error code (EC,Exx).
When the balance receives a control command:
The balance will send an AK code (acknowledgment, ASCII 06h) upon confirmation of receipt of the
command and completion of the process.
If the balance cannot execute the command, it sends an error code (EC, Exx).
The following control commands are handled by the balance, which sends an AK code
(acknowledgment, ASCII 06h) upon the completion of the process as well as upon confirmation of
receipt of the command.
If the balance cannot execute the command process, it sends an error code (EC, Exx). To clear the
error, use the CAL command.
Command
Description
ON
Turns the display on.
P
Turns the display on/off. (Only when the display is on.)
R, RZ
Same as the RE-ZERO button
.
T, TR
Same as the TARE button
.
ZR
Zero
If the load is within ±2% of the capacity from the initial zero point, the zero
point is updated, the tare value is cleared and the display is set to zero. If the
load exceeds ± 2%, no processing is done.
CAL
Sensitivity adjustment with the internal weight.
EXC
Sensitivity adjustment with an external weight.
TST
Performs calibration test with the internal weight.
RD:000
Closes the right door.*
RD:001
Opens the right door.*
LD:000
Closes the left door.*
LD:001
Opens the left door.*
* Supported with the touch panel software version 1.019 or later
Summary of Contents for BA-T Series
Page 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Page 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Page 117: ...117 Output example for periodic check output results Periodic check output results...
Page 118: ...118 Periodic check results PDF...
Page 129: ...129 AND MEET analysis graph example...
Page 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...