31
Normal weighing
Basic weighing
Example weighing in g for BA-6TE
Step 1. Press the ON:OFF key
or
the HOME key
to enter
weighing mode.
Step 2. Place a container (tare) on the weighing
pan if necessary. Press the RE-ZERO
button
to set the weighing value to
zero.
(The decimal separator position depends
on the balance model.)
Step 3. Place a sample on the pan or in the
container. Wait for the stabilization
indicator to be displayed. Read the value.
Step 4. Remove the sample and container
from the weighing pan.
ON:OFF key
or HOME key
Weighing pan
Container (tare)
RE-ZERO button
Zero display
Sample
Weighing
value
Zero after subtracting the tare
Step 1. If you press the ON:OFF key
with a container (tare)
placed on the weighing pan to display the weighing value,
zero is displayed after automatically subtracting the tare.
(Zero display after subtracting the tare)
Caution
In the zero display after subtracting the tare, the range
available for weighing becomes narrower than the
maximum display (capacity).
Range available for weighing = Maximum display - Tare
weight
For details about the range available for weighing up to the
capacity in the zero display during sensitivity adjustment when
the screen display is turned on, refer to “
Container (tare)
Weighing pan
ON:OFF key
Zero display
Summary of Contents for BA-T Series
Page 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Page 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Page 117: ...117 Output example for periodic check output results Periodic check output results...
Page 118: ...118 Periodic check results PDF...
Page 129: ...129 AND MEET analysis graph example...
Page 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...
Page 219: ......