Measurement example
Select a recipe to perform measurement according to the recipe. The registration example below is described
here as an example.
Recipe name:
Test
Samples to be
registered:
Sample name
Description
Target value
Tolerance
Sample1
Sample 1
1.000000 g
10.000 %
Sample2
Sample 2
1.000000 g
10.000 %
Step 1. On the [Formulation mode] screen, press the Go to the recipe list button to display the [Recipe list]
screen.
Select the recipe to be used for the measurement from the list and press the Back button to display the
[Formulation mode] screen.
Re-zero will be applied automatically when the screen changes to the [Formulation mode] screen.
Sample 1 is selected here as an example.
Step 2. To record the tare value:
Press the SAVE button
when the weighing value is 0 g or greater and stable.
Tare operation will be applied automatically when the SAVE button
is pressed.
The weighing screen for the first sample is displayed.
If the target value plus tare value of the first sample exceeds the capacity, the SAVE button
is
disabled.
To skip recoridng of the tare value:
Press the Skip button
.
The weighing screen for the first sample is displayed.
If a warning is displayed at the top of the screen:
Check the recipe name or the target value and tolerance value of the registered sample.
Step 3. Weigh the displayed sample.
When the stability indicator lights up and the SAVE button
is enabled, press the button to save the weighing value.
Allowable range: |(Weighing value) - (Target value)|
(Tolerance)
If the allowable range is exceeded:
To start over from tare weighing with the current sample, press
the Back button.
Step 4. Repeat the tare weighing and sample weighing in steps 2 and 3 for every sample registered in the recipe.
Step 5. The measurement results screen will be displayed when measurement of all samples is completed. Check
and output the results. Then, press the Back button to display the [Formulation mode] screen.
Summary of Contents for BA-T Series
Page 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Page 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Page 117: ...117 Output example for periodic check output results Periodic check output results...
Page 118: ...118 Periodic check results PDF...
Page 129: ...129 AND MEET analysis graph example...
Page 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...
Page 219: ......