1. Introduction
Thank you for purchasing A&D’s electronic balance.
Please read this instruction manual carefully to understand and make full use of the BA-T series
analytical electronic balance.
Caution
Operations may differ depending on the software version of your balance.
For confirmation of the software version of the balance, refer to “
Features
A 5-inch touch screen is adopted for easy operation.
Intuitive operation is provided through the touch screen.
Equipped with the breeze break auto doors that can be opened and closed without touching them.
The removable breeze break makes it easy to clean the inside of the weighing chamber.
A self check function is included for automatically evaluating repeatability performance without a weight.
The BA-T series can automatically perform sensitivity adjustment with the internal weight. (Automatic
sensitivity adjustment)
(Temperature change, set time, fixed time interval [interval time])
When performing the sensitivity adjustment/calibration test, etc. for the balance, the output
corresponding to GLP/GMP, etc. can be output.
Using a printer (sold separately), it is possible to record the sensitivity adjustment/calibration test
results.
GLP
:
Good Laboratory Practice. Standards for implementing safety tests for drugs and medicines.
GMP
:
Good Manufacturing Practice. Rules for manufacturing and quality control.
The clock built into the balance allows you to output the weighing value with the date and time.
(The clock settings can be restricted so that only the Administrator can change them. [Password
function])
The BA-T series comes standard with the underhook for weighing magnetic materials.
Six different applications are available.
Normal weighing, counting mode, percent weighing, minimum weight alert function, formulation mode,
and HPLC mode.
BA-6DTE, BA-225DTE and BA-125DTE have the smart range function as standard. This function allows
for weighing with the precision range after subtracting the tare within the weighing capacity.
The readability of precision range for BA-
6DTE is 1 μg.
The readability of precision range for BA-225DTE and BA-125DTE is
10 μg.
With the password function, the sensitivity adjustment of the balance and the operation of changing the
function table can be restricted.
An RS-232C interface, USB interface, Ethernet and Bluetooth for outputting the weighing value and
data of the balance are equipped as standard.
It is possible to communicate with a Windows personal computer that has data communication software
WinCT or WinCT-Plus installed. The software can be downloaded from our website
(
).
The BA-T series is equipped with a DC type ionizer (static eliminator) that does not generate wind, and
eliminates static electricity from charged objects before weighing in order to reduce errors due to static
electricity. The discharge electrode unit of the ionizer can be removed and can be cleaned and replaced
by itself.
Summary of Contents for BA-T Series
Page 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Page 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Page 117: ...117 Output example for periodic check output results Periodic check output results...
Page 118: ...118 Periodic check results PDF...
Page 129: ...129 AND MEET analysis graph example...
Page 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...
Page 219: ......